Model Number CLT-623-M
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/22/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was previously reported to the fda on a summary report and is now being submitted on a 3500a per fda request.Evaluation summary: post market vigilance (pmv) led an evaluation of received one devices.The visual inspection of the sample noted five needles.One needle was received broken at the tip.Upon microscopic inspection of the needle, instrument marks were observed on the needle adjacent to the break site.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: during the procedure and upon piercing the tissue with the needle, surgeon grasped the tip with a needle holder and tried to pull it up, the tip of the needle broke.The product was used for fascia suturing.There was no patient harm.
 
Additional Manufacturer Narrative
Initial notification of this reportable event to fda.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7316640
Report Number9612501-2018-00310
Device Sequence Number10800930
Product Code GAM
UDI-Device Identifier20884521044590
UDI-Public20884521044590
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model NumberCLT-623-M
Device Catalogue NumberCLT-623-M
Device Lot NumberD6J2184X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2017
Initial Date Received by Manufacturer 03/23/2017
Supplement Date Received by Manufacturer03/23/2017
Initial Report FDA Received Date03/06/2018
Supplement Report FDA Received Date03/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-