| Model Number |
8886640351 |
| Medical Device Problem Codes |
Bent (1059); Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event was previously submitted on a summary report.Subsequently, medtronic has received additional information which is being submitted via this 3500a.Post market vigilance (pmv) led an evaluation of three devices.The visual inspection and functional evaluation of the sample had acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic choledocholithotomy procedure, the needle broke during the closure of the incision.A new package was opened with new needle, but that needle was bent.The procedure was completed with another device with no issues.The patient did not suffer any injuries as a result of this complaint.
|
| |
|