• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number 8886640351
Medical Device Problem Codes Bent (1059); Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was previously submitted on a summary report.Subsequently, medtronic has received additional information which is being submitted via this 3500a.Post market vigilance (pmv) led an evaluation of three devices.The visual inspection and functional evaluation of the sample had acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic choledocholithotomy procedure, the needle broke during the closure of the incision.A new package was opened with new needle, but that needle was bent.The procedure was completed with another device with no issues.The patient did not suffer any injuries as a result of this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7316647
Report Number9612501-2018-00325
Device Sequence Number19233775
Product Code GAM
UDI-Device Identifier20884523003564
UDI-Public20884523003564
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model Number8886640351
Device Catalogue Number8886640351
Device Lot NumberD6J0699X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2017
Initial Report FDA Received Date03/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-