DAVIS & GECK CARIBE LTD STEEL; SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE Back to Search Results
Model Number 88862410-69
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/23/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was previously submitted on a summary report.Subsequently, medtronic has received additional information which is being submitted via this 3500a.Evaluation summary: post market vigilance (pmv) led an evaluation of four devices.The visual inspection and functional evaluation of the sample had acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: occurred during sternum closure.The device broke at the swage of the needle.Another device was opened to complete the procedure.No patient injury noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEEL
Common Device Name
SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7316650
Report Number9612501-2018-00324
Device Sequence Number11092036
Product Code GAQ
UDI-Device Identifier20884521102573
UDI-Public20884521102573
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K955723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 03/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model Number88862410-69
Device Catalogue Number88862410-69
Device Lot NumberD6K0755X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2017
Initial Report FDA Received Date03/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-