MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP Back to Search Results
Model Number 816571
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/09/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump had casting and bearing issues having a functional impact on the unit.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) observed the pump to function normally throughout the evaluation.
 
Manufacturer Narrative
The reported complaint could not be confirmed.The service repair technician (srt) was unable to duplicate the complaint as there were no issues with the castings or the bearings.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7317406
MDR Text Key101858392
Report Number1828100-2018-00109
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816571
Device Catalogue Number816571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2018
Was the Report Sent to FDA? No
Date Manufacturer Received09/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-