MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. VALEO BALLOON EXPANDABLE STENT; BILIARY STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. VALEO BALLOON EXPANDABLE STENT; BILIARY STENT Back to Search Results
Model Number EV07361CD
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the balloon expandable stent allegedly dislodged from the balloon before entering into the introducer sheath, rendering the device unusable.It was further reported that another balloon expandable stent was used to successfully perform the procedure.There was no patient contact.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.The device appeared to be clean.The stent was noted to be loose on the balloon and dislodge proximally from the distal marker band.No anomalies were noted to the stent.The balloon was examined under magnification and crimp marks were visible on the balloon.Therefore, based on the returned sample, the investigation is confirmed for dislodgement of the stent.The definitive root cause for the alleged dislodgement could not be determined based upon the available information.It is unknown if patient or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: warnings: should unusual resistance be felt at any time during the insertion process, do not force passage.If resistance occurs during movement through the sheath/guiding catheter, the stent/catheter should be withdrawn carefully.During implantation, if resistance occurs after the stent has exited the sheath/guiding catheter or if the stent cannot be delivered to the target location, attempts to retract the stent/catheter into the sheath/guiding catheter may result in dislodgement and embolization of the stent.The stent/catheter/sheath/guiding catheter should be removed as a single unit.Precautions: inspect the valeo biliary stent and delivery system prior to use to verify that the stent has not been damaged during shipment or handling and that the device dimensions are suitable for the specific procedure.Do not attempt to remove or adjust the stent on the delivery system.Stent/delivery system preparation: remove the stent/delivery system from its packaging [and remove the transport mandrel and balloon sheath from the distal end of the catheter if present].Inspect the stent for adherence to the balloon and centered placement in relation to the balloon marker bands.Do not reposition the stent or hand crimp.
 
Event or Problem Description
It was reported that the balloon expandable stent allegedly dislodged from the balloon before entering into the introducer sheath, rendering the device unusable.It was further reported that another balloon expandable stent was used to successfully perform the procedure.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALEO BALLOON EXPANDABLE STENT
Common Device Name
BILIARY STENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7317460
Report Number2020394-2018-00196
Device Sequence Number120141
Product Code FGE
UDI-Device Identifier00801741000751
UDI-Public(01)00801741000751
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K052132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2018
Device Model NumberEV07361CD
Device Catalogue NumberEV07361CD
Device Lot NumberGFAY0287
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Initial Date Received by Manufacturer 02/07/2018
Supplement Date Received by Manufacturer03/27/2018
Initial Report FDA Received Date03/06/2018
Supplement Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-