BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 014 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U4150215
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that while trying to load the pta balloon over the guidewire the distal tip of the balloon allegedly detached.Another balloon was used to perform the procedure.There was no patient contact.
 
Additional Manufacturer Narrative
Manufacturing review: the lot number has been provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported for this lot number and issue to date.Investigation summary: the device was returned for evaluation.A complete detachment of a portion of the distal tip was identified.The detached portion was not returned for evaluation.Therefore, the investigation is confirmed for the reported detached tip.The definitive root cause for the tip detachment could not be determined based on the available information.It is unknown if procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: dilatation catheter preparation: - remove the balloon guard and stylet by grasping the balloon catheter just proximal to the balloon and with the other hand, gently grasp the balloon protector and slide distally off of the balloon catheter.Precautions: - if resistance is felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.
 
Event or Problem Description
It was reported that while trying to load the pta balloon over the guidewire the distal tip of the balloon allegedly detached.Another balloon was used to perform the procedure.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 014 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7317646
Report Number2020394-2018-00198
Device Sequence Number10800906
Product Code LIT
UDI-Device Identifier00801741053443
UDI-Public(01)00801741053443
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121856
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model NumberU4150215
Device Catalogue NumberU4150215
Device Lot NumberCMBW0533
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Initial Date Received by Manufacturer 02/07/2018
Supplement Date Received by Manufacturer04/27/2018
Initial Report FDA Received Date03/06/2018
Supplement Report FDA Received Date05/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-