• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - ENGLEWOOD 1000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - ENGLEWOOD 1000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938739
Device Problem Disconnection (1171)
Patient Problem No Patient Involvement (2645)
Event Date 02/12/2018
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photograph revealed that the administration port cap was missing from the administration port.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the port cap fell off of the administration port of an exactamix 1000 ml eva tpn bag.This occurred when the nurse opened the overpouch prior to use.The bag had been compounded with pediatric parenteral nutrition and 7.5ml auspen trace elements.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7317781
MDR Text Key101730642
Report Number1416980-2018-01166
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2020
Device Catalogue NumberH938739
Device Lot Number1234180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-