MAUDE Adverse Event Report: BARD ACCESS SYSTEMS NAVARRE PERCUTANEOUS BILIARY DRAINAGE CATHETER WITH NITINOL, 8F LOCKING PIGTAIL; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS NAVARRE PERCUTANEOUS BILIARY DRAINAGE CATHETER WITH NITINOL, 8F LOCKING PIGTAIL; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number NBD8
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of gfbs1249 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that during preparation of insertion of the plastic stiffener, the user observed that the plastic stiffener was extremely fragile.The user tested the plastic stiffener before usage and it broke "as glass tube" in his hands without the use of great force.No patient involvement.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a damaged stylet was confirmed, but the exact cause was not determined.One open package with product code nbd8 and lot #gfbs1249 was returned for investigation.A broken stylet was found within the package.Three larger segments of stylet were found in the package.The proximal segment, which contained the white hub, was 14.5 cm in length.A second segment was 23cm in length.A third segment was 11.5cm in length.Four smaller fragments of the stylet were also found in the package.The smallest segment was 1.5mm x 1.0mm x 0.5mm.The fractured ends of the stylet segments had jagged edges.The product ifu cautions, ¿do not use excessive force when inserting stiffening cannula to straighten catheter (pigtail).Catheter tip may be straightened manually to aid in insertion of stiffening cannula.¿ the complaint sample was forwarded to the manufacturing facility for further review.A lot history review (lhr) of gfbs1249 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that during preparation of insertion of the plastic stiffener, the user observed that the plastic stiffener was extremely fragile.The user tested the plastic stiffener before usage and it broke "as glass tube" in his hands without the use of great force.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVARRE PERCUTANEOUS BILIARY DRAINAGE CATHETER WITH NITINOL, 8F LOCKING PIGTAIL
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7318007
Report Number3006260740-2018-00370
Device Sequence Number151268
Product Code FGE
UDI-Device Identifier00801741036866
UDI-Public(01)00801741036866
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
K955110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2020
Device Model NumberNBD8
Device Catalogue NumberNBD8
Device Lot NumberGFBS1249
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer03/23/2018
Initial Report FDA Received Date03/06/2018
Supplement Report FDA Received Date05/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-