Pump passed functional testings including displacement test, rewind test, basic occlusion test, occlusion test, prime/delivery test and excessive no delivery test.Drive support disk was inspected and no anomaly noted.Unit was received with normal operating currents and no unexpected off no power alarm or low battery alarm noted.Unit passed the dat test.Pump was received with cracked case at display window corner, minor scratched lcd window, cracked battery tube threads, cracked reservoir tube lip and stained address/serial number label.
|