MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SURGIDAC; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SURGIDAC; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 173026
Medical Device Problem Codes Break (1069); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: occurred during a laparoscopic hernia repair procedure.The suturing device was being used to suture very thick tissue.Once it was removed from the patient, it was difficult to unload the needle.It jammed and the needle broke outside of the patient.There is still a small piece of the needle in the jaw of the suturing device and the rest of the needle was retrieved from the floor.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one device.One broken needle piece was returned in the left jaw of suturing device.The other piece of the broken needle was received in a container.Needle pieces were inspected under a microscope and needle was observed to be broken at suture swage.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the bent or broken needle may occur when the needle is not loaded properly or when the needle is forced into an obstacle.This condition may also occur if excess pressure is exerted on the needle or the attached suture while the jaws are in the open position.In these situations, the needle may flex and ultimately break.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIDAC
Common Device Name
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7318166
Report Number9612501-2018-00387
Device Sequence Number104152
Product Code OCW
UDI-Device Identifier20884521129464
UDI-Public20884521129464
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K934738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number173026
Device Catalogue Number173026
Device Lot NumberJ7K1511X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2018
Initial Date Received by Manufacturer 02/15/2018
Supplement Date Received by Manufacturer04/05/2018
Initial Report FDA Received Date03/06/2018
Supplement Report FDA Received Date04/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-