| Model Number |
CC4204A |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).No similar complaint type events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
A cc4204a, +24.0 diopter, intraocular lens was returned damaged to the manufacturer.The reporter indicated that there was no patient contact with the lens.Additional information was unable to be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: device evaluation: the product was returned in liquid, in a lens case/vial.Visual inspection found the lens optic torn, the lens torn into two pieces, and missing a piece of haptic.Claim# (b)(4).
|
| |
|