JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1-PIECE; MONOFOCAL IOLS Back to Search Results
Model Number ZCB00
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable, as the lens was inserted and removed.If explanted, give date: not applicable, as the lens was inserted and removed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zcb00 15.5 diopter intraocular lens (iol) was noticed torn after being implanted in the patient's right eye (od).The lens was therefore removed and an incision enlargement was made with one suture being used.There was no patient injury and the patient is doing great post-operatively.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation, returned to manufacturer on: 03/08/2018.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection using microscope magnification was performed.The lens returned cut in two parts.Residue of viscoelastic was detected on both section of the lens optic and including the haptics.The lens condition observed is consistent with a unit that has been cut during surgical process.The reported issue was verified.However, based on the evidence observed the reported lens has not been affected by the manufacturing process.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation, there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7319373
Report Number2648035-2018-00341
Device Sequence Number10789214
Product Code HQL
UDI-Device Identifier05050474531628
UDI-Public(01)05050474531628(17)190820
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/20/2019
Device Model NumberZCB00
Device Catalogue NumberZCB0000155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2018
Initial Date Received by Manufacturer 02/06/2018
Supplement Date Received by Manufacturer03/13/2018
Initial Report FDA Received Date03/06/2018
Supplement Report FDA Received Date03/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-