Model Number PCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.If explanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.(b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the plunger could be moved normally during implantation.However, after implantation, when the intraocular lens (iol) opened / unfolded, there was a major defect on the iol surface/optic.The surgeon had to remove the iol and implant a new one.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation, returned to manufacturer on: 03/15/2018.Device evaluation: the sample was returned to the manufacturer.Pcb00 device delivery system was received with a sample container.The plunger was observed in a fully advanced position and locked.The cartridge was correctly engaged into the device.No assembly error and/or defect related to manufacturing process was observed.Visual inspection at 10x microscope magnification was performed and a lack of lubricant material was observed in the device.The pushrod was exposed out of the cartridge tip.The lens was not returned in or out of the device.Therefore, the reported issue could not be verified.The condition in which the sample was returned is consistent with a product that was handled and prepared for a surgical process.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.There was no discrepancy found during the review.A search revealed that no other complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7319431
Report Number2648035-2018-00343
Device Sequence Number18869816
Product Code HQL
UDI-Device Identifier05050474558137
UDI-Public(01)05050474558137(17)201207
Combination Product (Y/N)N
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/07/2020
Device Model NumberPCB00
Device Catalogue NumberPCB0000135
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2018
Initial Date Received by Manufacturer 02/06/2018
Supplement Date Received by Manufacturer03/23/2018
Initial Report FDA Received Date03/06/2018
Supplement Report FDA Received Date04/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-