• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN POLY GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN POLY GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 030775
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted that the stapler was received with a full complement of staples.The interlock was engaged.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the reported condition may occur if the firing knob is retracted prior to advancing.When the firing knob is retracted before firing the stapler it engages with the interlock in the handle and prevents the instrument from firing.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: during cystectomy/ileal conduit, the surgeon clamped the tissue of small intestine and tried to push the firing knob, however it was stiff and could not fire the device at all.Replaced by a new device, the procedure was completed.No injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLY GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7320076
Report Number1219930-2018-01240
Device Sequence Number105225
Product Code GDW
UDI-Device Identifier20884521047676
UDI-Public20884521047676
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K843603
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Model Number030775
Device Catalogue Number030775
Device Lot NumberN7C0362X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2017
Initial Report FDA Received Date03/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-