Model Number 7211010SR
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/01/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the transducer is not working since its overflowing.No patient injury or significant time delay reported.
 
Additional Manufacturer Narrative
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of over pressurization could not be reproduced.Product passed functional testing per factory test with no faults or errors.Product passed functional testing during 6 hour burn-in on wet station utilizing low and high pressure and flow settings.Raw and zero transducer readings were normal and well within specs during all functional tests.At no time during functional testing did pump over pressurize or become intermittent or fluctuate.All functions perform as expected.Pressure readings were normal throughout burn-in on wet station.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL {} C/U {} S&N DYONICS 25
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7320191
Report Number1643264-2018-00200
Device Sequence Number13480124
Product Code HRX
UDI-Device Identifier00885556586372
UDI-Public(01)00885556586372
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K171794
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7211010SR
Device Catalogue Number7211010SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2018
Initial Date Received by Manufacturer 03/01/2018
Supplement Date Received by Manufacturer04/17/2018
Initial Report FDA Received Date03/07/2018
Supplement Report FDA Received Date04/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-