| Model Number |
X SERIES |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during incoming inspection, the device failed to recognize the attached multifunction cable.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to zoll medical corporation.Instead, the suspect multi-function cable was returned.The malfunction was duplicated and attributed to a faulty multi-function cable.Analysis of reports of this type has not identified an increase in trend.
|
| |
|