| Catalog Number |
H1-LX |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician was attempting to use a hawkone directional atherectomy catheter to treat a severely calcified chronic total occlusion -100% in the mid left superficial femoral artery(sfa).The vessel was moderately tortuous.The device was prepped as per the ifu with no issues identified.The procedure used a 0.014" nitrex guidewire.It was reported that the wire split through the wire port during atherectomy.The procedure was completed by angioplasty.No patient injury was reported.
|
| |
|