• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT KYPHON NEUCHATEL MFG KYPHON XPANDER INFLATABLE BONE TAMP; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT KYPHON NEUCHATEL MFG KYPHON XPANDER INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number K09A
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Pre-operative diagnosis for this procedure: vertebral body compression fracture it was reported that intra-op, the balloon broke.The product came in contact with the patient.No patient symptoms or complications as a result of this event.
 
Additional Manufacturer Narrative
Product analysis: a visual and functional review reveal that the balloon has ruptured.The rupture is a pinhole near the proximal peak.This rupture is consistent with the balloon coming in contact with bone spurs with the balloon is inflated in the vertebral body.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MDT KYPHON NEUCHATEL MFG
pierre-à-bot 97
neuchâtel 2000
CH  2000
MDR Report Key7321075
Report Number2953769-2018-00012
Device Sequence Number109964
Product Code HRX
UDI-Device Identifier00858196001161
UDI-Public00858196001161
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberK09A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Initial Date Received by Manufacturer 02/05/2018
Supplement Date Received by Manufacturer05/01/2018
Initial Report FDA Received Date03/07/2018
Supplement Report FDA Received Date05/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-