It was reported that during the patient¿s generator replacement surgery, their explanted generator could not be interrogated.When their generator was replaced, a system diagnostic test was performed and high impedance observed.The lead pin was removed and reinserted with a second result of high impedance.Therefore the lead was inspected, and the surgeon noticed a small tear in the lead.The lead was therefore explanted and a new lead was implanted.Additional information was received from the surgeon¿s office stating that the tear in the lead was present in the outer and inner tubing.The surgeon¿s believed cause of the reported tear in the lead was stated to be due to the patient falling 6 months prior when they sustained a left clavicle fracture.Lastly it was stated that the vns has been working poorly for those 6 months.The explanted products have not been received by the manufacturer to date.No other relevant information has been received to date.
|
The lead and generator were received by the manufacturer for product analysis.The lead underwent product analysis and abraded openings were observed on the outer silicone tubing with a small abraded opening on one of the inner silicone tubes.Several fractures were observed in the returned portion of the lead.Scanning electron microscopy was performed which identified pitting and residual material on the fractured pieces.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.The abraded openings found on the inner and outer silicone tubing and the cut ends that were made during the explanted process, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer and inner silicone tubing.The set screw marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.With the exception of the observed discontinuity and abraded tubing openings, the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The generator underwent product analysis and the battery status was confirmed to be set at eos (end of service) = yes.With the exception of the end of service battery status setting, the device performed according to functional specifications.Analysis concluded proper functionality of the pulse generator and that no abnormal performance or any other type of adverse condition was found.No other relevant information has been received to date.
|