| Model Number |
MICL12.6 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 12.6mm micl12.6 implantable collamer lens, -10.0 diopter, in the patient's right eye (od) on (b)(6) 2018.The lens was explanted on (b)(6) 2018 due to excessive vaulting.The lens was exchanged for a shorter lens and the problem was resolved.The reporter indicated the cause of the event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the lens was returned dry, in a (b)(4) bag.Visual inspection found a piece of one haptic torn off and missing.There was evidence of clear residue on the lens.(b)(4).
|
| |
|