Model Number 3971
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Date of Event 01/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of one device.The single use loading unit (sulu) was received partially applied.The sulu was loaded into a pmv test unit.The instrument and sulu were applied to the appropriate test media with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during sub total gastrectomy, they opened the package and tried to connect cartridge to device, however they noticed the staples of the proximal end came out of the staple pockets.They stopped using this cartridge and replaced by the new cartridge, now the firing is complete.The event occurred during the procedure but the product was not used for patient.Patient there was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7322061
Report Number2647580-2018-01245
Device Sequence Number120219
Product Code GDW
UDI-Device Identifier20884521063164
UDI-Public20884521063164
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K885222
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number3971
Device Catalogue Number3971
Device Lot NumberP7E0389X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2018
Initial Report FDA Received Date03/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-