| Model Number |
3971 |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.The single use loading unit (sulu) was received partially applied.The sulu was loaded into a pmv test unit.The instrument and sulu were applied to the appropriate test media with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during sub total gastrectomy, they opened the package and tried to connect cartridge to device, however they noticed the staples of the proximal end came out of the staple pockets.They stopped using this cartridge and replaced by the new cartridge, now the firing is complete.The event occurred during the procedure but the product was not used for patient.Patient there was no patient injury.
|
| |
|
Search Alerts/Recalls
|