| Model Number |
515-019000F |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Failure to Sense (1559)
|
| Health Effect - Clinical Code |
Misdiagnosis (2159)
|
| Date of Event |
01/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The system was removed from service and the amplifier was returned to natus for evaluation.The evaluation found no problems with the amplifier.A natus field service technician is going on site to evaluate the system.
|
| |
|
Event or Problem Description
|
|
Eeg brain test; waves appeared flat, in reality they were not flat.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up report for 3010611950-2018-00001.Reference (b)(4) for complaint, (b)(4) for onsite information.Mdr follow up: field service went on site and evaluated the system and found that the headbox had intermittent inputs.The headbox was replaced.A clinical representative reviewed the onsite information and stated that in these types of situations, eeg is not the only factor considered.Many other tests would be conducted including a thorough neuro exam.
|
| |
|
Event or Problem Description
|
|
Eeg brain test; waves appeared flat, in reality they were not flat.
|
| |
|