• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); Visual Disturbances (2140)
Date of Event 04/26/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/date of birth: unknown, information not provided.If explanted, give date: not applicable, as the lens remains implanted.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zlb00 20.0 diopter intraocular lens (iol) was implanted in the patient's operative eye on (b)(6) 2017.Post-operatively, since the day after surgery, the patient has been experiencing worse near and distance vision and a ''ghosting line,'' which has not gone away.Reportedly, the patient has been back to her surgeon a number of times looking for some guidance.The patient is currently using glasses, which has improved her vision, but it is still not as good as the patient would like.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to the specifications.A search revealed that no similar complaints were received from this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7322414
Report Number9614546-2018-00230
Device Sequence Number10793017
Product Code MFK
UDI-Device Identifier05050474560673
UDI-Public(01)05050474560673(17)200729
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date07/29/2020
Device Model NumberZLB00
Device Catalogue NumberZLB00U0200
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/08/2018
Supplement Date Received by Manufacturer04/10/2018
Initial Report FDA Received Date03/07/2018
Supplement Report FDA Received Date05/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-