| Model Number |
CQ2015A |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Collamer ultraviolet-absorbing posterior chamber three piece foldable intraocular lens.Work order search: lens was returned in liquid, in lens case vial.Visual inspection found the lens torn in two pieces and portion of the optic torn (missing).Work order search: one similar complaint type event was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter stated that the surgeon was inserting a cq2015a implantable collamer lens, +21.0diopter, in the patients right eye (od) and a haptic tore off.There was patient contact but no injury.The lens was removed and the back up lens was implanted within the same surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|