• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number 1013465-200
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported balloon rupture.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during a peripheral procedure, the 3.0 x 200 mm armada 18 dilatation catheter was advanced to the target lesion in the mildly calcified anterior tibial artery.The armada 18 balloon was inflated to 8 atmospheres and ruptured during the first inflation.The device was removed without difficulty.There was no adverse patient effect and no clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7322453
Report Number2024168-2018-01649
Device Sequence Number145272
Product Code LIT
UDI-Device Identifier08717648207938
UDI-Public08717648207938
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number1013465-200
Device Lot Number7092541
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2018
Initial Report FDA Received Date03/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
Patient Weight93
-
-