| Catalog Number |
TEXP140RUH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
01/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6)-year-old caucasian female patient underwent a primary breast reconstruction procedure with a mentor artoura breast tissue expander 850cc that deflated after implantation.The tissue expander stopped filling and there was a tear on the seam of the device.The patient fully recovered and she experienced no accompanying symptoms.As a result of the deflation, the patient underwent explantation and replacement with a mentor artoura breast tissue expander 850cc on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
On 05/09/2018, mentor completed the analysis on the suspect medical device.Upon receipt by mentor, the device returned without fluid.No foreign material was observed within the device or on the shell surface.The product evaluation (pe) team discovered a rent at the juncture with the patch.Microscopic examination of the edges of the rent gave no indications as to cause.No other anomalies were discovered.The complaint was confirmed since a rupture was found on the device.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet; however, deflation trends for mentor saline-filled devices will continue to be monitored by quality assurance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|