Device was used for treatment, not diagnosis.It is unknown when non-union occurred and device failure occurred.Event description updated.Implant date: unknown date in (b)(6) 2017.It is unknown if devices were explanted and if they were explanted dates are unknown.Device is not expected to return for investigation.Pma #: provided.Device history record review, manufacturing location: (b)(4), manufacturing date: 29-sep-2016, expiration date: 31-aug-2025.Part #: 04.003.460s, lot#: h196945 (sterile) ¿ 11 mm ti lateral entry femoral recon nail-ex/400 mm/rt - sterile.Quantity 6.Inspection sheet met inspection acceptance criteria.Component parts reviewed: part 21014, lot: 7922124 meet specifications.Raw material for titanium was received from dynamet incorporated.Product certification was received from dynamet and raw material receiving/putaway checklist meet specification.(b)(4) was initiated at operation 190 for boxing on (b)(6) 2016.This is a planned non conformance for "(b)(4) rev.1 does not require to repeat steps 3.08 and 3.09 for each part." (b)(4) was initiated at operation 20 for gundrill and 40 bend radius on (b)(6) 2016 for "implementing the jbs for gun drill in the nail department, six pack holders are called out on (b)(4) these are still awaiting production due to machine times." issue is not relevant to the complaint condition.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.(b)(4), ethicon (abq), ¿sterility documentation was reviewed and determined to be conforming.¿6: device history record review the investigation could not be completed; no conclusion could be drawn, as no product was received.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
|