Catalog Number 71442011
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/21/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the tip broke during surgery.No reported injury to the patient.
 
Additional Manufacturer Narrative
The associated complaint device was returned for evaluation.A visual inspection of the returned r3 offset impactor revealed moderate signs of usage and wear.The reported issue is confirmed as part of the tip of the device on the posterior end has sheared of which may cause malfunction.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This investigation could not verify or identify any evidence of product contribution to the reported problem.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.We consider this investigation closed.Credit will be issued for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN CONE IMPACTOR HANDLE
Common Device Name
PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAIND,CMNTD,POLYMR/METAL/POLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7323167
Report Number1020279-2018-00283
Device Sequence Number1605220
Product Code JWH
UDI-Device Identifier00885556537152
UDI-Public00885556537152
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K162775
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71442011
Device Lot Number17ASC0011
Initial Date Received by Manufacturer 02/21/2018
Supplement Date Received by Manufacturer02/21/2018
Initial Report FDA Received Date03/08/2018
Supplement Report FDA Received Date07/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-