| Model Number |
8450107035D |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the bur broke due to issues identified in the associated attachment (5407120950a ;sn (b)(4)).Corrosion was identified in the attachment.The attachment front bearing overheated and developed wear on the bur weakening the neck strength and potentially causing the bur to break.The quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
|
| |
|