| Catalog Number |
AB18W060100150 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Codes |
Patient Problem/Medical Problem (2688); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician was attempting to use a powercross balloon to treat a severely calcified and slightly tortuous mid left iliac artery lesion exhibiting 100% occlusion.The device was used with a non-mdt sheath <(>&<)> guidewire.No embolic protection was used.The device was removed from its packaging without issue and inspected with no issues noted.The ifu was followed during the procedure.The device did not pass through a previously deployed stent.It was reported that resistance was encountered when advancing the device and that excessive force was used during delivery of the device.During the initial balloon inflation, using a syringe and contrast/saline mix, a balloon burst occurred.Following this, it was reported that a fragment of the balloon remained in the patient.The patient attended the or where the balloon fragment was removed by common femoral endarterectomy.No further patient injury was reported.
|
| |
|
Search Alerts/Recalls
|