| Model Number |
LF1520 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during nephrectomy, the device worked fine in the beginning but then the jaws got locked onto tissue and did not open.The jaws were somehow forcibly opened and released from tissue, but then the jaws did not open and close smoothly.New device was used to continue.No patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.The jaws opened and closed properly when the handle was retracted and released.The knife blade advanced and retracted smoothly along the knife track.The knife cut of the device was tested on a silicone test strip with acceptable results.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|