| Model Number |
M00537510 |
| Medical Device Problem Code |
Device Packaging Compromised (2916)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complainant indicated that the device will not be returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.If any further relevant information is received, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an advanix¿ pancreatic stent was to be used for an unknown procedure on an unknown date.According to the complainant, during unpacking, it was observed that the packaging of the stent was already damaged and the sterile seal was compromised.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual evaluation of the returned device found there was cut on the tyvek side of the package.The cut appears to has been made with a sharp tool.Based on the product analysis, the failure was noted during unpacking and outside the patient, it is possible that the way in which the device was handled and manipulated during its process of unpacking may have contributed to the condition of the unit returned (package cut).The damage found was caused with an unknown/sharp instrument since the section of the package found cut has its appearance without irregularities and the cut looks very clean.During its manufacturing process the product is inspected in regards to its integrity, however there is no control on how the units are handled at the field, it was concluded the way in which the unit was handled and manipulated during preparation may have contributed to the failure.Therefore, ¿handling damage¿ is selected as the most probable root cause for the complaint.A dhr (device history record) review was performed and no deviation was found.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an advanix¿ pancreatic stent was to be used for an unknown procedure on an unknown date.According to the complainant, during unpacking, it was observed that the packaging of the stent was already damaged and the sterile seal was compromised.There were no patient complications reported as a result of this event.
|
| |
|