| Model Number |
7209161 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tipe broke due to excessive force.All parts were removed.No significant delay or patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Examination was not possible, as the device has not been returned.The investigation could not draw any conclusions about the reported event without the return of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
One 180 degree cuff stitch left was returned for evaluation.Device is over four years old.Visual assessment confirmed the reported breakage.The distal tip at beginning of first bend has been broken off.The shaft is bent.Examination of the break area shows evidence of abnormal twisting/bend at fracture point.The condition of the device indicates it was subjected to excessive forces during use.
|
| |
|