| Model Number |
H1-M |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
02/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Image analysis: seven photographic images of the device were received for evaluation five of the photographic images are of the distal segment of the distal assembly.The guidewire is through the distal tip and appears to be torn from the guidewire lumen over the coil section of the distal assembly.Two photographic images are of the distal segment and proximal segment of the distal assembly along with the cutter/drive shaft.The distal assembly separated just distal of the cutter window where the coil segment of the distal assembly starts.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician intended to use a hawkone 67f, 7f sheath and 6.0 spider fx, to treat a moderately tortuous, calcified lesion, with 60-70% stenosis, in the proximal, mid and distal superficial femoral artery (5-6mm diameter),.The device was prepared as per ifu.It was reported that the physician encountered resistance while removing the hawkone and the tip of the device detached.A vascular cut down was performed to remove the tip of the hawkone from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Hawkone was returned with inserted cutter driver.There was no damage to the torque shaft.The cutter assembly and drive shaft were intact and pushed forward.The fracture of the distal assembly occurred distal the vent holes.The fracture shows the as the laser drilled coils star, they are stretched out at the area of the fracture.The damage to the h1 is consistent with a prolapse event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|