| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); No Code Available (3191)
|
| Date of Event |
01/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that he implanted a 13.2mm micl13.2, -4.5 diopter, implantable collamer lens in the patient left eye (os) on (b)(6) 2018.On post op day 1 excessive vault, elevated iop and angle closure was discovered.The patient was placed on combigan to lower iop.At day 3 and 5 the elevated vault persisted with narrowing of the angle.On (b)(6) 2018 the lens was exchanged for a shorter lens and the problem was resolved.The reported stated that the patient is happy with good current vision.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: device evaluation: lens was returned in liquid in a lens vial.Visual inspection found haptic tears.This is a re-submission to correct mfr # for supplemental #1 in which the mfr number was reported incorrect.(b)(4).
|
| |
|