| Catalog Number |
03P81-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat ec4+ cartridges that yielded a suspected discrepant hematocrit & hemoglobin results on a patient.There was no patient information available at the time of this report.Return product is not available.(b)(6).Sample a repeated 30 minutes later.Test and collect times:unk.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).The investigation was completed on 03/15/2018.A review of the device history record confirmed the lot passed finished goods release criteria.Retain cartridge test results met the acceptance criteria found in q04.01.003 rev.Ac, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency identified.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|
Search Alerts/Recalls
|