MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS INSERT SZ 3-4 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS INSERT SZ 3-4 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 71420818
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 02/08/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a total knee replacement, the insert was not able to effectively mount on baseplate.The surgery delayed 30 min and bleeding increased and blood transfusions were needed.
 
Additional Manufacturer Narrative
The associated complaint device was returned and evaluated.A visual inspection of the returned insert indicated damage to the locking detail and the dove tail.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A dimensional inspection was attempted; the damage/deformation at several features of the device would not allow for accurate measurement.The features that could be measured were within specification.A clinical analysis indicated that based on the information provided, the inability to ¿mount¿ the 11mm pe insert was the contributing factor to the 30 minute surgical delay; however, per the product evaluation with dimensional inspection, the root cause of the failure to mount could not be determined.The patient impact was the 30 minute surgical delay which reportedly did not result in any harm or injury to the patient.No further medical assessment is warranted at this time.A review of complaint history on the listed part revealed no additional complaints for the listed batch.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS INSERT SZ 3-4 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7325459
Report Number1020279-2018-00288
Device Sequence Number151298
Product Code JWH
UDI-Device Identifier03596010209580
UDI-Public03596010209580
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71420818
Device Lot Number17CT83702
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2018
Initial Date Received by Manufacturer 02/09/2018
Supplement Date Received by Manufacturer02/09/2018
Initial Report FDA Received Date03/08/2018
Supplement Report FDA Received Date04/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient Weight58
-
-