|
Patient height reported as 162.0 cms.Date of event: date of non-union is not known.Additional product codes: hrs, ktt.Date of implant/therapy date reported as (b)(6) 2017, exact date is not known.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that a patient was implanted with a 3.5mm locking compression plate (lcp) posterolateral distal humerus plate, 3.5mm cortex screws (quantity of 3), 2.7mm locking screws (quantity of 2) and 3.5mm locking screws (quantity of 4) for an orif (open reduction internal fixation) of a humerus in (b)(6) 2017.A nonunion was noted on an unknown date.X-rays were taken on an unknown date and showed that the (two) 2.7mm locking screws were not connected to the plate.Both screws were stuck in the bone fully intact, causing minimal fixation.The x-rays also showed that the threads on one of the 2.7mm locking screw heads stripped.Patient underwent revision surgery on (b)(6) 2018.All hardware was easily removed.Patient was revised to a variable angle (va) plate, screws and vivigen formable bone grafts.There was no surgical delay.The procedure was completed successfully and the patient was reported as stable.Concomitant devices reported: 3.5mm cortex screw, self-tapping 20mm (part 204.820, lot number unknown, quantity 1); 3.5mm cortex screw, self-tapping 24mm (part 204.824 lot number unknown, quantity 1); 3.5mm cortex screw, self-tapping 28mm (part 204.828 lot number unknown, quantity 1); 3.5mm locking screw, self-tapping with stardrive recess 22mm (part 212.107 lot number unknown, quantity 2) and 3.5mm locking screw, self-tapping with stardrive recess 26mm (part 212.109 lot number unknown, quantity 2).This report is for one (1) 2.7mm locking screw.This is report 3 of 3 for pc-(b)(4).
|