• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number DR8044
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/26/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during treatment for a shunt, the pta balloon allegedly leaked on the first inflation.Reportedly, there was no issue retracting the balloon through the sheath.Another balloon was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.An initial visual inspection did not find any signs of leaking.However, once an inflation attempt was made, a leak and partial break were noted at the butt joint between the balloon and catheter shaft.Therefore, the investigation is confirmed for the reported leak, as well as for a break at the butt joint.The partial break at the glue butt joint resulted in the identified leak.However, the definitive root cause for the partial break could not be determined based upon available information.It is unknown whether procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: if resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Use of the dorado pta dilatation catheter: advance the catheter through the introducer sheath and over the wire to the site of inflation.If the stenosis cannot be crossed with the desired dilatation catheter, use a smaller diameter catheter to pre-dilate the lesion to facilitate passage of a more appropriately sized dilatation catheter.Position the balloon relative to the lesion to be dilated, ensure the guidewire is in place and inflate the balloon to the appropriate pressure.
 
Event or Problem Description
It was reported that during treatment for a shunt, the pta balloon allegedly leaked on the first inflation.Reportedly, there was no issue retracting the balloon through the sheath.Another balloon was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7325629
Report Number2020394-2018-00210
Device Sequence Number13854513
Product Code DQY
UDI-Device Identifier00801741058301
UDI-Public(01)00801741058301
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2020
Device Catalogue NumberDR8044
Device Lot Number93NB0245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Initial Date Received by Manufacturer 02/23/2018
Supplement Date Received by Manufacturer03/20/2018
Initial Report FDA Received Date03/08/2018
Supplement Report FDA Received Date03/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-