• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number 1013470-040
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual and functional inspection was performed on the returned device.The reported leak was confirmed.Additionally, a crack in the hub was noted.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that could have contributed to this event.Based on review of the similar incidents, there is no indication of a lot specific issue.The investigation determined the reported leak was due to the noted cracks on the hub and appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that while backloading the armada 18 balloon dilatation catheter (bdc) onto the guide wire, negative pressure was applied and air bubbles were noted at the proximal end.The device was removed and not used in the anatomy.A different armada 35 bdc was used successfully in the procedure, without issue.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7325661
Report Number2024168-2018-01678
Device Sequence Number11132019
Product Code LIT
UDI-Device Identifier08717648208393
UDI-Public08717648208393
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Catalogue Number1013470-040
Device Lot Number6100541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2018
Initial Report FDA Received Date03/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient Weight54
-
-