| Model Number |
ZA9003 |
| Medical Device Problem Code |
Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
11/15/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Unknown if the lens was implanted.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection at 10x microscope magnification showed residue of lubricant material and loose particles on the lens surface indicating that the product was handled and prepared for surgical process.Both haptics were observed distorted/bent.Some scratches were observed on the lens surface.The customer's reported complaint was verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue was verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an intraocular lens (iol) was scratched.No further information was provided.
|
| |
|