Catalog Number 02.07.0214FUC
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 09 march 2018; lot 080861: (b)(4) items manufactured and released on 02 june 2008 expiration date: 2013-03-31 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Clinical evaluation performed by medical affairs on 09 march 2018: insert revision in tka 7,5 years after primary for joint dislocation.The insert was exchanged to a thicker one.The developmental instability is a rather commonly described possible complication following tka, because soft tissue may stretch and provide insufficient stability.It is therefore common practice to resort to a thicker insert and recreate soft tissue tension.This is not a problem due to any implant defect or malfunction.
 
Event or Problem Description
The patient came in due to dislocation.The surgeon revised the 14mm insert with a 17mm insert.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-PRIMARY TIBIAL INSERT UC FIXED SIZE 2 / 14 MM
Common Device Name
TIBIAL INSERT UC FIXED
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key7326624
Report Number3005180920-2018-00128
Device Sequence Number18859371
Product Code JWH
UDI-Device Identifier07630030816659
UDI-Public07630030816659
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2013
Device Catalogue Number02.07.0214FUC
Device Lot Number080861
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2018
Initial Report FDA Received Date03/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-