Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/09/2018
Event Type  malfunction  
Manufacturer Narrative
A gehc service representative performed a checkout of the equipment and confirmed the reported complaint.Inspection of the equipment noted that the elbow near the regulator had popped off.The elbow was reseated.Patient information could not be obtained after multiple attempts.Attempts were made as follows: 3/5/18 phone call, 3/6/18 phone call, 3/7/18 phone call.
 
Event Description
The hospital reported that, during a case, a leak was heard.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7326699
MDR Text Key102091343
Report Number2112667-2018-00515
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042154
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/09/2018
Initial Date FDA Received03/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-