| Model Number |
STB-30C4S |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stabilizer was returned in a biohazard bag inside the original shipper box with the label attached.Functional evaluation of the device was performed.The device did not show any defect in this testing.The stabilizer was mounted on a retractor rail.The base locked in place and the stabilizer arm tightened with the handle.The arm held as expected and no defects were found.Chase medical was unable to recreate the failure during functional evaluation.Accordingly, it appears the problem was user error.Review of the lot history record showed nothing out of the ordinary.All testing was performed and documented with no issues seen during the manufacture of the lot.Functional testing upon return of the device could not duplicate the reported failure and the device appeared fully functional.
|
| |
|
Event or Problem Description
|
|
As described - "arm wouldn't hold on stabilizer during surgery".There was no harm to the patient and they just used another stabilizer.Chase medical was notified of this complaint on (b)(6) 2018 by customer, (b)(6).This incident occurred at (b)(6) general hospital.
|
| |
|