• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE THORACIC; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE THORACIC; PLATE Back to Search Results
Model Number 24-039-30-72
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Impaired Healing (2378); Patient Problem/Medical Problem (2688)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Reference exemption number e2017029.
 
Event or Problem Description
It was reported that the patient had an off mid-line sternotomy in which the ribs were separated from the sternum at the coastal condylar cartilage.The doctor fixated the lss plates to the ribs in a vertical fashion, which is not the method for which they are indicated.Ifu states "plates are not designed for transverse fracture fixation, rib-to-rib fixation, or the spanning of continuity defects." according to the doctor the lower portion of the plate placed on the ribs was removed due to a patient infection unrelated to the product.
 
Additional Manufacturer Narrative
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage falls within the design risk limits adhered to at klm.The device history records could not be reviewed due to the lot number not being provided by the customer.The root cause for the failure cannot be determined due to no device being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE THORACIC
Common Device Name
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
MDR Report Key7327265
Report Number9610905-2018-00035
Device Sequence Number10789380
Product Code HRS
UDI-Device Identifier10888118097924
UDI-Public(01)10888118097924
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K151983
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/02/2018
Report Date (Section F) 03/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number24-039-30-72
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Hospital
Date Report to Manufacturer03/02/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2018
Supplement Date Received by Manufacturer04/06/2018
Initial Report FDA Received Date03/09/2018
Supplement Report FDA Received Date04/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-