BECTON, DICKINSON & CO., (BD) BD VACUTAINER® WINGED SAFETY PUSH BUTTON BLOOD COLLECTION SET
|
Back to Search Results |
|
Catalog Number 367342 |
Device Problems
Break (1069); Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/22/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medical device expiration date: unknown.Device manufacture date: unknown.Investigation: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Additionally, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.As no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
|
Event Description
|
It was reported that the safety of a bd vacutainer® winged safety push button blood collection set ¿broke off¿ during use.There was no report of exposure, injury or medical intervention.
|
|
Manufacturer Narrative
|
Bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for a broken and missing iv needle retraction button with the incident lot was observed.Additionally, the samples were observed to be heavily damaged, the iv needles were retracted and the samples were missing the male luer.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Based on evaluation of the customer samples, the customer¿s indicated failure mode for a broken and missing iv needle retraction button with the incident lot was observed.Based on the investigation, a root cause could not be determined.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
|
Search Alerts/Recalls
|
|
|