| Model Number |
MMT-7020B |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 92 mg/dl and the sensor glucose was 61 mg/dl at the time of the incident.The customer's blood glucose level was 95 mg/dl.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected 1 opened/used sensor and performed continuity resistance test.Sensor passed per specification.Also, performed bicarbonate buffer test.Sensor passed per specifications with accurate readings.See attached file for details.Also found sensor with cannula bent.Unable to confirm customer received sensor in said condition due to customer returned opened/used.
|
| |
|