| Catalog Number |
ZNGRMS180HS |
| Medical Device Problem Codes |
Component or Accessory Incompatibility (2897); Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Intimal Dissection (1333); Death (1802); Thrombus (2101)
|
| Date of Event |
02/07/2018
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician intended to use a zinger guide wire in a procedure to treat a moderately tortuous, non-calcified lesion in the proximal - mid right coronary artery (rca) with 80% proximal stenosis and 99% mid stenosis.The patient had no previous anterior pci.The patient had no previous stents, or other products implanted.The patient had a critical obstruction in the right coronary vessel caused by atheroma.There was no damage noted to the device packaging.The device was removed from the packaging per ifu with no issues.The device was inspected with no issues.The device was prepped per ifu with no issues.The wire tip was not formed by the physician.The lesion was not pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device and excessive force was not used.It is reported that there was low support, poor steerability/poor navigability during delivery to the target lesion and deformation of the guidewire.Other devices were in use with the wire when the issue occurred, a la6jr40 launcher guide catheter, a pyton kit and a 6f introducer (061102a).There is no evidence about the use of an angioplastic balloon for support in the vessel.A dissection occurred in the proximal segment of the right coronary artery.The physician implanted an integrity bare metal stent (int25018x) in the dissection site.The patient experienced sub-acute thrombosis and patient death occurred on same day as the procedure.The physician reports that the incident is related to the guidewire.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the wire exhibited kinks 7.5cm, 102cm, 122cm, 126cm from the proximal end of the wire.A spiral curl of about 4 turns was noted 9cm from the proximal coils joint.Severe stretching was noted between the wire first joint and the proximal coils joint.The distal wire ribbon between the first joint and the proximal coils joint was broken approximately 2mm from the first joint.The distal tip remained attached by the coils to the remaining proximal wire.The whole length of the wire was returned, no parts of the wire are missing.Actual wire outer diameter passed physical measurements, as per specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|