| Catalog Number |
21-7023-24 |
| Medical Device Problem Code |
Occlusion Within Device (1423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).This device is same and/or similar to a device approved for distribution in the us.
|
| |
|
Event or Problem Description
|
|
It was reported that a cadd® administration set caused the pump to alarm "upstream occlusion".This occurred during use after the set-up of the cassette and the infusion tube, which was noted as being "very difficult".The alarm was cleared after replacing the infusion tube.No injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One cadd® cadd® administration set was returned in used condition.The sample was visually inspected at a distance of 12" to 24" under normal conditions of illumination.The join between the star tube and the pump tube was detached from the housing.No marks of solvent were found in this part of the product.No obstructions were detected in any of the joins.The manufacturing and quality inspection processes were reviewed and considered adequate and correct.The failure mode could not be duplicated.The most probable root cause is that the tubing detached activated the sensors on the pump for an occlusion alarm.
|
| |
|