MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C2860-A
Medical Device Problem Codes Kinked (1339); Device Damaged Prior to Use (2284); Device Handling Problem (3265)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The ruby coil pusher assembly was kinked approximately 97.0, 100.0 and 101.0 cm from its proximal end.Evaluation of the returned ruby coil revealed kinks in close proximity on the pusher assembly.These kinks were likely due to forcefully gripping and advancing the coil against resistance.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
While preparing a ruby coil for a coil embolization procedure, a hospital staff inadvertently kinked the ruby coil pusher assembly while loading the coil into a lantern delivery microcatheter (lantern) on the back table.The kinking of the ruby coil occurred prior to use and therefore, the ruby coil was not used in the procedure.The procedure was completed using a new ruby coil and the same lantern.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7328525
Report Number3005168196-2018-00530
Device Sequence Number109942
Product Code HCG
UDI-Device Identifier00814548013091
UDI-Public00814548013091
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2025
Device Catalogue NumberRBY2C2860-A
Device Lot NumberF78562
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2018
Initial Report FDA Received Date03/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-